Mercury Analyzers,Mercury Instruments,Mercury Analysis,Mercury Vapor Monitor,Mercury Testing,Hg Analysis
LA-254  
Products Home PageAULA-254ASDVM-3000Mercury Tracker            3000PA-1SM-3MMSUT-3000MC-3000Which Instrument?Technology

  Img6.jpgView Manual

Optimization of AAS technology

In contrast to a typical multielement AAS the LabAnalyzer 254 is specially designed for elemental mercury. This allows top performance in analytical applications. The use of a specially developed highly stable mercury lamp in connection with thermostat-controlled UV sensors results in a detection limit of a few ppt of mercury.

Memory effects are minimized and a high sample throughput is possible thanks to specially selected materials for sample gas conducting components and heating of the optical bench.


Operation and maintenance
The LabAnalyzers 254 is operated with a waterproof membrane keypad and a graphic LCD. Handling is extremely easy. The LabAnalyzer 254 does not require auxiliary gases. The carrier gas flow required for the analysis is generated by a built-in membrane pump which is maintenance-free and has a long service life.

Measurement of a sample
The sample is pipetted into an extractor and mixed with 0.5 ml of reducing agent, for example tin-II-chloride solution. The extractor is then connected to the LabAnalyzer 254 and the measurement is started by pressing a button. After 60 to 90 sec the result appears on the display and the device is ready for the next analysis. The system does not need to be purged, and the zero point is calibrated automatically before every measurement (auto zero).
 

Safety for the user
The mercury cannot escape into the working environment, as any free mercury is collected in a sulfurized activated carbon cartridge. If the cartridge needs replacement a message appears on the control panel. The analyzed sample no longer contains any mercury.

Short analyses
Even with high concentrations no lengthy purging time is required. After every measurement a zero calibration is carried out automatically (auto zero). The typical duration of a measurement including purging is 60 to 100 sec. over the entire measuring range.

Display of results
During the entire measurement the measuring signal is continuously displayed on the graphic display. An acoustic alarm is sounded to indicate the end of the measurement. In addition to the measurement signal curve over time, the peak value and the mercury concentration are indicated.

Storage of analytical results and quality assurance
Together with the data necessary for quality assurance the results of the analyses are stored in the RAM of the built-in computer. They are available at any time, and an automatic print-out of this data is also possible.
 

LA-254 LabAnalyzer Technical Specifications
Measuring principle: UV-Absorption
Wavelength: 254 nm
UV source: electrodeless low-pressure mercury lamp (EDL), temperature-controlled
UV detectors: Silicon, UV-enhanced, temperature controlled
Stabilization method: Double beam (reference beam) method
Optical cell: Fused silica (Suprasil), approx. 23 cm long
Optical cell heating: approx. 70°C
Measuring range: lowest range: 0...1 µg/L
highest range: 0...100 mg/l
Pump: long-life membrane pump
Air stream: 30 l/h, adjustable
Measuring range: 0.01 µg/l ... 10 µg/l (10 ppt ... 10 ppb) for 10 ml
Sensitivity: 5 ng/l or 0.05 ng
Sample volume: 2 ... 10 ml
Reducing agent: tin-II-chloride or sodiumborohydride
Display of readings: Graphic LCD with background illumination
Outputs: analog 4-20 mA, RS 232computer output, parallel Centronics printer output
Power Supply: 230 VAC/ 50-60 Hz (optional 115 VAC / 50-60Hz)
Power consumption: 35 W
Dimensions: 45 x 15 x 35 cm (Wx H x D) photometer section
24 x 48 x 27 cm (Wx H x D) reaction bottle section
Required floor space: approx. 70 x 50 cm (W x D)
Weight: approx. 10 kg

 

 

Products | Home Page | AULA-254 | ASD - Automatic Sample Digestion | VM-3000 | Mercury Tracker 3000 | PA-1 | SM-3 | MMS | Ultratracer UT-3000 | MC-3000 Mercury Calibrator | Which Instrument? | Technology